Studi Quality Control (QC) Pesawat Fluoroscopy (Angiografi) di PT. Siloam Internasional Hospital Makassar menggunakan Multimeter RaySafe (X2) dan Black Piranha RTI
DOI:
https://doi.org/10.26418/positron.v8i2.24752Keywords:
Quality control, Fluoroscopy, AngiografiAbstract
Tujuan dari penelitian ini adalah melakukan uji quality control (QC) pada pesawat fluoroscopy (angiografi) di PT. Siloam Internasional Hospital Makassar, pengukuran didasarkan pada protokol QC yang dikembangkan oleh Radiological Council Western Australia. Pengukuran yang dilakukan meliputi akurasi tegangan tabung (kVp), waktu fluoroskopik maksimum, dan uji kualitas berkas/half value layer (HVL), linearitas keluaran (mGy/mAs), laju dosis permukaan II, dan kualitas citra. Hasil penelitian menunjukkan bahwa pengujian generator dan tabung pesawat sinar-x masih dalam batas toleransi yaitu 1,4%. Nilai maksimum laju dosis pasien tipikal 14,3 mGy/min, dan laju dosis serap reseptor citra sebesar 67,8 mGy/min untuk pengukuran menggunakan 20 cm phantom polymethyl methacrylate (PMMA) yang dioperasikan pada mode dosis normal dan dosis tinggi.
References
Marshall, N.W., Lecomber, A.R., Kotre, C.J., and Faulkner K., Fluoroscopy quality assurance measurements : automatic exposure rate control and image quality, Radiat. Prot. Dosim., 80(1-3), pp 69-72, 1998.
Suliman, I.I., van Soldt, R.T., and Zoetelief, J., Digital fluoroscopy quality control measurement, In World Congress on Medical Physics and Biomedical Engineering 2006, pp 1470-1474, 2007.
Glaiberman, C., How to create a quality assurance program for radiation safety in interventional radiology, Tech. Vasc. Interv. Radiol., 3(3), pp 194-199, 2010.
Walsh, C., Dowling, A., Meade, A., and Malone, J., Subjective and objective measure of image quality in digital fluoroscopy, Radiat. Prot. Dosimetry, 117(1-3), pp 35-37, 2012.
Dimov, A. and Vassileva, J., Quality conrol of the x-ray fluoroscopy equipment in the period of their clinical use - methods and first results, 9 National Conference on Biomedical Physics and Engineering, 2004.
Radiation Safety ACT, Diagnostic x-ray equipment compliance testing, 1975.
Leeds Test Objects, Fluoro 4 user manual, pp 3 - 6, 2011.
Downloads
Published
Issue
Section
License
Authors who publish with this journal agree to the following terms:
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).

This work is licensed under a Creative Commons Attribution 4.0 International License.